QUANTIFICAÇÃO DE ANTIBIÓTICOS BETA-LACTÂMICOS EM MEDICAMENTOS VETERINÁRIOS: DETERMINAÇÃO DE AMOXICILINA E AMPICILINA POR CROMATOGRAFIA LÍQUIDA DE ALTA EFICIÊNCIA

2013 
This work focused on the development and validation of an RP-HPLC-UV method for quantification of beta-lactam antibiotics in three pharmaceutical samples. Active principles analyzed were amoxicillin and ampicillin, in 3 veterinary drugs. Mobile phase comprised 5 mmol L-1 phosphoric acid solution at pH 2.00, acetonitrile with gradient elution mode and detection wavelength at 220 nm. The method was validated according to the Brazilian National Health Surveillance regulation, where linear range and linearity, selectivity, precision, accuracy and ruggedness were evaluated. Inter day precision and accuracy for pharmaceutical samples 1, 2 and 3 were: 1.43 and 1.43%; 4.71 and 3.74%; 2.72 and 1.72%, respectively, while regression coefficients for analytical curves exceeded 0.99. The method had acceptable merit figure values, indicating reliable quantification. Analyzed samples had active principle concentrations varying from -12 to +21% compared to manufacturer label claims, rendering the medicine unsafe for administration to animals.