Application of high intensity focused ultrasound for treatment of open-angle glaucoma in Indian patients

2018 
Purpose: The aim of this study is to assess the efficacy of ultrasound cycloplasty (UCP) in Indian patients with open angle glaucoma (OAG). Methods: A prospective interventional study was designed to treat 73 eyes of 73 patients with OAG with the EyeOP1 device equipped with six miniaturized cylindrical piezoelectric transducers. Two treatment protocols of ultrasound delivery depending on exposure time (8 s and 10 s of shot per transducer) were used. Complete ophthalmic examination, ocular biometry and anterior segment optical coherence tomography were performed preoperatively and patients were followed up at day 1, day 7 and months 1, 2, 3, 6, and 12. The primary outcome measure was “successful” intraocular pressure (IOP) control defined as IOP reduction ≥20% from baseline and IOP value >5 mm Hg at the last follow-up visit. Secondary outcomes were the occurrence of complications and mean IOP during the follow-up period. Results: In all patients, the mean IOP reduced from 23.5 ± 3.0 mmHg before treatment to 15.7 ± 5.4 mmHg at 12 months ( P Conclusion: Our short-term results reveal that UCP is a simple, safe, and noninvasive procedure which enables to significantly reduce the IOP in patients with OAG. The study results in Indian eyes corroborate findings in earlier studies on Caucasian eyes.