Safety of Electromagnetic Articulography in Patients With Pacemakers and Implantable Cardioverter-Defibrillators

2009 
Purpose Electromagnetic articulography (EMA) uses a helmet to create alternating magnetic fields for tracking speech articulator movement. An important safety consideration is whether EMA magnetic fields interfere with the operation of speakers' pacemakers or implantable cardioverter-defibrillators (ICDs). In this investigation, individuals with pacemaker/ICD devices were exposed to EMA fields under controlled conditions while potential interference was examined. Method Twelve adults with pacemaker/ICD devices from 3 major manufacturers were assessed for device function before, during, and after exposure to magnetic fields from a Carstens AG100 EMA system. Potential interference was probed, with EMA transmitters positioned at varying distances from the implantable devices and with the EMA system set at different operating strengths. Results No adverse affects in device operation were observed under any conditions. The only potential complication was temporary telemetry-link interference during device test...