Small balloon strategy associated with low pacemaker implantation rate after self-expanding transcatheter valve implantation.

2021 
BACKGROUND This study aims to investigate whether small balloon aortic valvuloplasty (BAV) reduces the need for permanent pacemaker implantation (PPMI) after transcatheter aortic valve implantation (TAVI). METHODS This was a retrospective analysis using data from our local TAVI database. Small BAV was defined as a small balloon size (=18 mm) pre-dilatation. Normal BAV was defined as a balloon size >18 mm. The primary endpoint was the incidence of new PPMI. RESULTS Of 99 consecutive TAVI patients, five patients were excluded due to pre-existing permanent pacemaker. Patients in the small BAV group (n=57) had a significantly lower PPMI rate compared with the normal BAV group (n=37) (3.5% vs. 18.9%, P=0.026). Moderate or severe aortic valve regurgitation post-procedure was similar between the small BAV and normal BAV groups (5.3% vs. 8.1%, P=0.480); likewise, the mean aortic gradient post-procedure did not differ significantly (11.5±5.2 mmHg vs. 12.2±7.3 mmHg, 1 mmHg=0.133 kPa, P=0.580) between the groups. Device success rates were also similar (94.7% vs. 91.8%, P=0.680). In multivariable analysis, small BAV (P=0.027), the ratio of prosthesis diameter to annulus diameter (P=0.048), and mean aortic gradient by echo in the basement (P=0.021) were independent predictors of PPMI. CONCLUSIONS The small BAV strategy is associated with a low rate of permanent pacemaker implantation after transcatheter self-expanding valve implantation in this single-center observational study.