Ex vivo Biomechanical Comparison of V‐Loc 180® Absorbable Wound Closure Device and Standard Polyglyconate Suture for Diaphragmatic Herniorrhaphy in a Canine Model

2015 
OBJECTIVE: To compare 2-0 polyglyconate barbed suture (V-LOC™ 180 Absorbable Wound Closure Device, Covidien, Mansfield, MA) to standard 2-0 monofilament absorbable polyglyconate suture (Maxon™, Covidien) for maximum load to failure and failure mode when used to appose standardized defects in canine diaphragm muscle. STUDY DESIGN: Randomized, cadaveric ex vivo mechanical testing. SAMPLE POPULATION: Canine cadaveric hemi-diaphragmatic specimens (n = 32 pairs). METHODS: Thirty-two cadaveric diaphragm specimens were collected and divided to obtain 64 muscle specimens (9 cm × 4 cm). Paired specimens were bisected in the midpoint of their 9 cm length in the direction of the muscle fibers and apposed using either 2-0 polyglyconate or 2-0 barbed polyglyconate (V-LOC™ 180 Absorbable Wound Closure Device) in a simple continuous pattern. Increasing tension was applied perpendicular to the suture line until specimen failure. Failure mode and maximum load to failure were recorded for each specimen. RESULTS: Mean ± SD failure load of simulated herniorrhaphies performed with barbed polyglyconate suture (54.5 ± 10.27 N) was not significantly different than repair with polyglyconate (56.9 ± 10.87 N). Failure mode for both construct types was suture tear out rather than suture failure. CONCLUSIONS: Clinical application of 2-0 barbed polyglyconate suture can be considered as an alternative to the use of 2-0 polyglyconate for diaphragmatic herniorrhaphy.