New Detector Sensitivity Calibration and the Calculation of the Interaction Force between Particles Using an Optical Tweezer

2018 
We propose a new approach to calculate the sensitivity factor of the detector in optical tweezers. In this work, we used a charge-coupled device (CCD) camera and a quadrant photodiode (QPD) for the extraction of the various positions occupied by the trapped object (in this case, silica beads of different diameters). Image-J software and the Boltzmann statistical method were then used to estimate the sensitivity of the detector. Silica beads of diameter 0.8 µm, 2 µm, a system of 2 µm bead stuck to 4.5 µm one and another system of 2 µm beads stuck to 2 µm one, were studied. This work contributes significantly to making better calibration of the detector without taking into account the geometry of the object imprisoned in the optical trap. We further developed an approach to calculate the interaction force between two microbeads. This approach does not require any knowledge of solvent viscosity and works for all types of samples.