Abstract Beam tracking as a mitigation technique for treatment of intra-fractionally moving organs requires prediction to overcome latencies in the adaptation process. We implemented and experimentally tested a prediction method for scanned carbon beam tracking. Beam tracking parameters, i.e. the shift of the Bragg peak position in 3D, are determined prior to treatment in 4D treatment planning and applied during treatment delivery in dependence on the motion state of the target as well as on the scanning spot in the target. Hence, prediction is required for the organ motion trajectory as well as the scanning progress to achieve maximal performance. Prediction algorithms to determine beam displacements that overcome these latencies were implemented. Prediction times of 25 ms for target spot prediction were required for ∼6 mm water-equivalent longitudinal beam shifts. The experimental tests proved feasibility of the implemented prediction algorithm.