Time-related changes in neointimal tissue coverage of a novel Sirolimus eluting stent: Serial observations with optical coherence tomography.

2016 
Abstract Background DES has reduced rates of restenosis compared with BMS but it has been associated with delayed healing and increase of stent thrombosis. The aim of our study was to evaluate the vascular time-related changes following implantation of a new SES coated with an amorphous silicon carbide that allows faster re-endothelisation (Orsiro-Biotronik). Methods This prospective registry enrolled STEMI-patients with multi-vessel disease, thus candidates for a two-step procedure. PCI of the culprit lesion was performed with at least one Orsiro stent that was OCT-analysed during the second-step procedure (deferred to 30,90 and 180-days). Results 16 of the 95 patients with MVD underwent an OCT evaluation of the study device implanted in the culprit lesion during the second-step staged procedure and were enrolled in the present Registry. A total of 3060-struts were analysed. The percentage of uncovered struts was 19.6% at30-days, 1.3% at90-days and 1.8% at180-days (p  Conclusions Our data shows that the new Orsiro stent promotes early and persistent strut coverage with low peri-strut thrombus. This pilot OCT evaluation might suggest a low incidence of late adverse events and anticipate safe outcome after early withdrawal of dual antiplatelet therapy.