Lung nodules : Dual-kilovolt peak analysis with CT-multicenter study

2000 
PURPOSE: To test the following hypothesis: The greater the increase in the mean computed tomographic (CT) number of a radiologically indeterminate lung nodule from the CT number on a 140-kVp CT image to that on an 80-kVp CT image, the more likely the nodule is benign (ie, contains calcium). MATERIALS AND METHODS: Two hundred forty indeterminate lung nodules were prospectively studied at four institutions: Mayo Clinic Scottsdale, Ariz (n = 160); Mayo Clinic Rochester, Minn (n = 50); Shiga Health Insurance Hospital, Otsu, Japan (n = 25); and Duke University Medical Center, Durham, NC (n = 5). Of the 240 nodules, 157 met the entrance criteria for this study and had a diagnosis. All nodules included were solid, 5–40-mm diameter, relatively spherical, homogeneous, and without visible evidence of calcification or fat. Each nodule was evaluated by using 3-mm-collimation, nonenhanced CT scans with both 140- and 80-kVp x-ray beams. RESULTS: There were 86 (55%) benign and 71 (45%) malignant nodules. The median incr...