False‐Positive Behavior with the dP/dt Sensing Pacemaker: A Rare Complication of a Physiological Sensor

1997 
As with “nonphysiological” devices, sensors that directly measure physiological variables bave the potential to measure unexpected signals and for the physiological parameter being measured to respond in an unexpected manner. We present tbe case of a dP/dt sensing pacing system that functioned normally for 2 months and then developed upper rate behavior due to the sensing of a high frequency artifact on the pressure recording. Our case and others cited reinforce the need for future physiological rate responsive pacemakers to incorporate a second sensor to provide for backup rate response in cases of inappropriate rate response.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    5
    References
    2
    Citations
    NaN
    KQI
    []