Determine scattering coefficient and anisotropy of scattering of murine tissues using reflectance-mode confocal microscopy
2013
Different techniques have been developed to determine the optical properties of turbid media, which include
collimated transmission, diffuse reflectance, adding-doubling and goniometry. While goniometry can be used to
determine the anisotropy of scattering (g), other techniques are used to measure the absorption coefficient and
reduced scattering coefficient (μs(1-g)). But separating scattering coefficient (μs) and anisotropy of scattering
from reduced scattering coefficient has been tricky. We developed an algorithm to determine anisotropy of
scattering from the depth dependent decay of reflectance-mode confocal scanning laser microscopy (rCSLM)
data. This report presents the testing of the algorithm on tissue phantoms with different anisotropies (g = 0.127
to 0.868, at 488 nm wavelength). Tissue phantoms were made from polystyrene microspheres (6 sizes 0.1-0.5
μm dia.) dispersed in both aqueous solutions and agarose gels. Three dimensional images were captured. The
rCSLM-signal followed an exponential decay as a function of depth of the focal volume, R(z)ρexp(-μz) where
ρ (dimensionless, ρ = 1 for a mirror) is the local reflectivity and μ [cm-1] is the exponential decay constant.
The theory was developed to uniquely map the experimentally determined μ and ρ into the optical scattering
properties μs and g. The values of μs and g depend on the composition and microstructure of tissues, and allow
characterization of a tissue.