Dose profile measurement using an imaging plate: Evaluation of filters using Monte Carlo simulation of 4 MV x-rays

2009 
Computed radiography (CR) is gradually replacing film. The application of CR for two-dimensional profiles and off-axis ratio (OAR) measurement using an imaging plate (IP) in a CR system is currently under discussion. However, a well known problem for IPs in dosimetry is that they use high atomic number (Z) materials, such as Ba, which have an energy dependency in a photon interaction. Although there are some reports that it is possible to compensate for the energy dependency with metal filters, the appropriate thicknesses of these filters and where they should be located have not been investigated. The purpose of this study is to find the most suitable filter for use with an IP as a dosimetric tool. Monte Carlo simulation (Geant4 8.1) was used to determine the filter to minimize the measurement error in OAR measurements of 4 MV x-rays. In this simulation, the material and thickness of the filter and distance between the IP and the filter were varied to determine most suitable filter conditions that gave t...