Respiratory inductive plethysmography accuracy at varying PEEP levels and degrees of acute lung injury

2006 
Background and objective: This study was performed to assess the accuracy of respiratory inductive plethysmographic (RIP) estimated lung volume changes at varying positive end-expiratory pressures (PEEP) during different degrees of acute respiratory failure.Methods: Measurements of inspiratory tidal volume were validated in eight piglets during constant volume ventilation at incremental and decremental PEEP levels and with increasing severity of pulmonary injury. RIP accuracy was assessed with calibration from the healthy state, from the disease state as the measurement error was assessed, and at various PEEP levels.Results: Best results (bias 3%, precision 7%) were obtained in healthy animals. RIP accuracy decreased with progressing degrees of acute respiratory failure and was PEEP dependent, unless RIP was calibrated again. When calibration was performed in the disease state as the measurement error was assessed, bias was reduced but precision did not improve (bias – 2%, precision 9%).Conclusions: RIP a...