Design and Testing of the NCTU Micro-Orifice Cascade Impactor (NMCI) for the Measurement of Nanoparticle Size Distributions

2015 
The NCTU micro-orifice cascade impactor (NMCI) was redesigned and tested to enable the measurement and adjustment of the jet-to-plate distance (S) of the impactors by using a micrometer. Each stage of the NMCI contains an impaction plate and a nozzle plate which are separated. The bottom casing for impaction plate assembly and the nozzle plate holder of each stage are separated. Calibration results show that the cutoff aerodynamic diameter (dpa50) of the three lower stages (dpa50 = 180, 100, and 56 nm) are close to the nominal values given in Marple et al. (1991). In addition, the S/W (W: nozzle diameter) effects on the dpa50 for the three lower stages were investigated and an empirical equation was developed to facilitate the prediction of dpa50 when the nozzle diameters may change slightly from one batch to another. The relationship between the nozzle diameter and pressure drop of the micro-orifice impactors at the fixed operational flow rate of 30 L/min was also established so that the nozzle diameters...