Clinical implementation of respiratory‐gated spot‐scanning proton therapy: An efficiency analysis of active motion management

2019 
Purpose The aim of this work is to describe the clinical implementation of respiratory‐gated spot‐scanning proton therapy (SSPT) for the treatment of thoracic and abdominal moving targets. The experience of our institution is summarized, from initial acceptance and commissioning tests to the development of standard clinical operating procedures for simulation, motion assessment, motion mitigation, treatment planning, and gated SSPT treatment delivery.