Performance Evaluation of Serum Free Light Chain AnalysisNephelometry vs Turbidimetry, Monoclonal vs Polyclonal Reagents

2017 
Objectives: Free light chain (FLC) measurement gained a lot of interest for diagnostic workup of monoclonal gammopathy. Methods: We evaluated the performance of turbidimetric polyclonal Freelite (The Binding Site, Birmingham, UK) assays on Cobas 6000 (Roche Diagnostics, Rotkreuz, Switzerland) and nephelometric monoclonal N Latex (Siemens Healthcare Diagnostics, Marburg, Germany) assays on BN ProSpec (Dade Behring, Deerfield, IL) vs established nephelometric Freelite assays on BN ProSpec. Results: Analytical performance was acceptable. Method comparison (n = 118) showed significant proportional FLC differences for N Latex assays. However, good correlation and clinical concordance were shown. Recovery study in the low concentration range demonstrated consistent over- and underrecovery for Freelite reagents, hampering future research on prognostic value of suppressed noninvolved FLC. Antigen excess detection was successful for κ FLC in three-fourths of cases with Freelite reagents and in all cases with N Latex reagents. However, the latter resulted in underestimated κ FLC concentrations. Conclusions: FLC analysis requires continuous awareness of analytical limitations. Monitoring of disease response requires FLC analysis on the same platform using the same reagents.