First Measurements with New High-Resolution Gadolinium-GEM Neutron Detectors
2016
European Spallation Source instruments like the macromolecular diffractometer (NMX) require an excellent neutron detection efficiency, high-rate capabilities, time resolution, and an unprecedented spatial resolution in the order of a few hundred micrometers over a wide angular range of the incoming neutrons. For these instruments solid converters in combination with Micro Pattern Gaseous Detectors (MPGDs) are a promising option. A GEM detector with gadolinium converter was tested on a cold neutron beam at the IFE research reactor in Norway. The mTPC analysis, proven to improve the spatial resolution in the case of 10 B converters, is ex- tended to gadolinium based detectors. For the first time, a Gd-GEM was successfully operated to detect neutrons with a measured efficiency of 12% at a wavelength of 2 A and a position resolution better than 400 mm.