Determination of Carbon Monoxide in Tuna by Gas Chromatography with Micro-Thermal Conductivity Detector

2008 
The suitability of a portable gas chromatograph equipped with a micro-thermal conductivity detector for the head-space determination of carbon monoxide (CO) in tuna samples is evaluated; CO is estimated after its liberation from tissue by acidic treatment at 70°C. Using the tested technique, the CO contents in untreated and suspected treated samples are analyzed. A limit of detection of approximately 13 ng/g is reached. The results demonstrate that this apparatus has performances similar to more expensive and sophisticated instruments.