QUADRICEPS EFFICIENCY FOLLOWING TOTAL KNEE ARTHROPLASTY WITH A MEDIAL-PIVOT SYSTEM IS SIMILAR TO CONTROL

2018 
IntroductionQuadriceps performance following total knee arthroplasty (TKA) is a critical factor in patient satisfaction that can be significantly affected by implant design (Greene, 2008). The objective of this study was to compare quadriceps efficiency (QE) following TKA with a medial-pivot system (EVOLUTION®, MicroPort Orthopedics Inc., Arlington, TN, USA) to non-implanted control measurements.MethodsFive cadaveric leg specimens with no prior surgeries, deformities, or disease were obtained. Each was placed in a custom closed chain device and loaded to simulate a heel-up squat from full-extension to deep flexion (approximately 115°) and back to full extension. Quadriceps force (FQ) and ground reaction force (FZ) were measured, and the ratio of the two was calculated as the quadriceps load factor (QLF). QFLs are inversely related to QE, with higher QFLs representing reduced efficiency. Each specimen was then implanted with a medial-pivot implant by a board certified orthopedic surgeon and force measureme...