Disposable Strip-type Sensors for Detection of Free Chlorine

2012 
The measurement of residual chlorine as a disinfectant is very important to ensure the safety against the pathogenic microbes and to suppress injection. The portable free chlorine sensor was fabricated with a disposable strip format by a screen printing method. The strip sensors prepared with a carbon-Ag/AgCl(cathode-anode) combination exhibited less interfering responses towards combined chlorine species(especially NHCl2) and oxygen than the sensors prepared with other metals(i.e., gold and platinum). Free chlorine was determined chronoamper- ometrically with carbon-based electrodes at an applied potential of -0.3 V(vs. Ag/AgCl). A channel was built on the strip-type electrode for easy sampling, and the resulting strip sensors were employed to determine the concentrations of residual free chlorine.