Towards in-vivo K-edge imaging using a new semi-analytical calibration method
2014
Flat field calibration methods are commonly used in computed tomography (CT) to correct for system imperfections.
Unfortunately, they cannot be applied in energy-resolving CT when using bow-tie filters owing to spectral distortions
imprinted by the filter. This work presents a novel semi-analytical calibration method for photon-counting spectral CT
systems, which is applicable with a bow-tie filter in place and efficiently compensates pile-up effects at fourfold
increased photon flux compared to a previously published method without degradation of image quality. The achieved
reduction of the scan time enabled the first K-edge imaging in-vivo. The method employs a calibration measurement with
a set of flat sheets of only a single absorber material and utilizes an analytical model to predict the expected photon
counts, taking into account factors such as x-ray spectrum and detector response. From the ratios of the measured x-ray
intensities and the corresponding simulated photon counts, a look-up table is generated. By use of this look-up table,
measured photon-counts can be corrected yielding data in line with the analytical model. The corrected data show low
pixel-to-pixel variations and pile-up effects are mitigated. Consequently, operations like material decomposition based
on the same analytical model yield accurate results. The method was validated on a experimental spectral CT system
equipped with a bow-tie filter in a phantom experiment and an in-vivo animal study. The level of artifacts in the resulting
images is considerably lower than in images generated with a previously published method. First in-vivo K-edge images
of a rabbit selectively depict vessel occlusion by an ytterbium-based thermoresponsive polymer.