Glyphosate immunosensor. Application for water and soil analysis.

2005 
A fully automated immunosensor for the herbicide glyphosate has been developed on the basis of the immunocomplex capture assay protocol. The sensor carries out on-line analyte derivatization prior to the assay and uses a selective anti-glyphosate serum, a glyphosate peroxidase enzyme tracer, and fluorescent detection. Under optimal conditions, the detection limit achieved is 0.021 μg/L with an analysis rate of 25 min per assay, autonomy of more than 48 h, and sensor reusability >500 analytical cycles. The immunosensor is able to discriminate structurally related molecules, such as aminomethylphosphonic acid, the main metabolite of glyphosate, and other related herbicides, such as glufosinate and glyphosine. Interferences from naturally occurring species (anions, cations, and humic substances) and their elimination were also studied. The immunosensor has been successfully applied to water and soil sample analysis, with good recoveries at levels lower than 1 μg/L. Results obtained with the immunosensor corr...