Use of Excess Solid-Phase Capacity in Immunoassays: Advantages for Semicontinuous, Near-Real-Time Measurements and for Analysis of Matrix Effects

2004 
A flow-based immunoassay system using solid-phase particles with high binding capacity was used for semicontinuous, near-real-time, measurement of 17β-estradiol (E2). The high binding capacity of the solid phase was exploited to enable (i) a quantitative determination of E2 concentration, based on rate of accumulation of fluorescently labeled anti-E2 antibody on the solid phase, and (ii) the use of a single solid phase for more than a dozen competitive binding measurements. The high binding capacity of the solid phase also permitted the immobilization of a second capture antigen. Biotin was immobilized as a second antigen and used to evaluate a biotin anti-biotin system as a control for matrix effects in the E2 immunoassay. In phosphate-buffered saline, E2 could be quantified (in the range of 10−1000 pM) by using either the summation or ratio of the signals from the labeled anti-E2 and anti-biotin antibody in the presence of biotin at a constant concentration. The same referencing system was applied to es...