Determination of cadmium, zinc, and lead by non-dispersive atomic fluorescence spectrometry with a new graphite furnace atomizer

1976 
Abstract A new graphite furnace atomizer has been developed and applied to the determination of cadmium, zinc, and lead by non-dispersive atomic fluorescence spectrometry. A solar-blind photomultiplier, a lock-in amplifier, and microwave-excited electrodeless discharge lamps are used. The detection limits for cadmium, zinc, and lead in the non-dispersive atomic fluorescence mode are 1·10 −13 g, 2·10 −13 g, and 2·10 −11 g, respectively, which are 20-, 10-, and 2-fold better than those in the atomic absorption mode. The analytical working curves are linear over about three decades of concentration from the detection limits.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    15
    References
    5
    Citations
    NaN
    KQI
    []