On the β-detection efficiency of a combined Si and plastic stack detector for DESPEC

2020 
Abstract A Geant4 simulation has been carried out in order to determine the β -detection efficiency of a rare isotope beam implantation setup, for decay spectroscopy experiments, comprising a number of Double Sided Silicon Strip Detectors (DSSSDs) and two plastic scintillation detectors placed upstream and downstream. The absolute efficiency for the emitted β -particle detection from radioactive fragments implanted in the DSSSDs using fast-timing plastic-scintillator detector, is calculated. The detection efficiency of the setup has been studied with two different distances between the Si layers and plastics. The requirement for the thickness of the Si detector layers and its implication on the β -detection efficiency has been investigated for 1 mm and 300 μ m thickness of Si layers. The combined efficiency of DSSSD and plastic detectors were also simulated for two different thicknesses of the DSSSD.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    1
    Citations
    NaN
    KQI
    []