Device interaction between cardiac contractility modulation (CCM) and subcutaneous defibrillator (S-ICD).

2021 
Combined implantation of cardiac contractility modulation (CCM) with subcutaneous implantable cardioverter-defibrillator (S-ICD) appears a suitable option to reduce the amount of intracardiac leads and complications for patients. Here we report on a patient with ischemic cardiomyopathy carrying an S-ICD in which a CCM device was implanted. During crosstalk testing post-CCM implantation, the S-ICD misannotated QRS complexes and T waves. The problem was solved through reprogramming the CCM, while preserving S-ICD functionality and improving heart failure symptoms. In conclusion, S-ICD combined with CCM seems to be a good and safe option for patients when device interference is being ruled out.