Absolute, soft x-ray calorimetry on the Z facility at Sandia National Laboratories (invited)

1999 
Simple and reliable x-ray fluence measurements, in addition to time-resolved diagnostics, are needed to understand the physics of hot Z-pinch plasmas. A commercially available laser calorimeter has been modified for measuring soft x-ray fluence from the Z facility at Sandia National Laboratories. The x-ray absorber of this calorimeter is an aluminum disk, attached to a two-dimensional thermopile and surrounded by an isoperibol shroud. The time integral and the maximum of the thermopile voltage signal are both proportional to the x-ray energy deposited. Data are collected for 90 s, and the instrument has, thus far, been used in the 1–25 mJ range. A wider dynamic measuring range for x-ray fluence (energy/area) can be achieved by varying the area of the defining aperture. The calorimeter is calibrated by an electrical substitution method. Calibrations are performed before and after each x-ray experiment on the Z facility. The calibration of the time integral of the thermopile voltage versus energy deposited ...