Comparative Dose of Contrast-Enhanced Spectral Mammography (CESM), Digital Mammography, and Digital Breast Tomosynthesis

2018 
OBJECTIVE. The purpose of this study is to provide a more accurate estimation of the radiation dose of contrast-enhanced spectral mammography (CESM) relative to that of 2D digital mammography and tomosynthesis using phantom and patient data and an accepted dosimetry protocol that eliminates vendor-specific average glandular dose (AGD) estimates while including breast density. MATERIALS AND METHODS. Patient and phantom AGD estimation was performed using two vendors (system 1 and system 2) in five imaging modes, including 2D, 3D, and CESM imaging. Patient AGD was retrospectively estimated from 45 patients who underwent mammography with all imaging modes during 2012–2016. Patient and phantom AGD were estimated using accepted European and International Atomic Energy Agency protocols for dosimetry and were compared across imaging modes using a paired t test with Bonferroni correction. RESULTS. Phantom data showed that the imaging modes with the lowest to highest AGDs were system 1 2D, followed by system 2 2D a...