A study on sensitivity correction factor of gamma camera in whole body bone image

2014 
AbstractThe aim of this study was to calculate a reasonable correction factor that could be corrected when the sensitivity of each gamma camera is measured and a whole body bone scan is performed as a follow-up examination. Seven pieces of equipment were used to analyse the sensitivity measurements of each gamma camera based on the source sensitivity measurement method recommended by the IAEA. These were BRIGHTVIEW, PRECEDENCE, ECAM, ECAM signature, ECAM Plus, SYMBIA T2 and INFINIA. The 99mTc line source for the sensitivity measurement was generated in 4–7 kcps, which are the common values in a whole body bone scan. All the cameras were equipped with a low-energy and high-resolution parallel multi-hole collimator and set at a window width and photo peak of 15% and 140 keV, respectively. After placing the 99mTc source as closely as possible to the collimator, the count was measured for 60 and 120 seconds to calculate correction factor. To determine if the correction factor calculated using the 99mTc line s...