Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010.
2021
OBJECTIVE The aim of this study was to determine the accuracy of the PHILIPS DL8760, an oscillometric blood pressure (BP) measuring device designed for self-measurement in the general population according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. PARTICIPANTS AND METHODS The PHILIPS DL8760 measures BP at the brachial level. The validation was performed in 33 subjects. The mean age of the participants was 55.8 ± 14 years. SBP 146.1 ± 26 mmHg (range: 103-193), DBP was 86.8 ± 17.0 mmHg (range: 47-123), and arm circumference was 28.3 ± 4.0 cm (range: 22-39). RESULTS The PHILIPS DL8760 fulfilled the criteria of the ESH-IP by passing phases one and two for both SBP measurements and DBP measurements. The device overestimated SBP by 1.5 ± 5.0 mmHg and underestimated DBP by 0.7 ± 3.9 mmHg. CONCLUSION The PHILIPS DL8760 BP device, with its specially designed cuff covering a broad range of arm circumferences, fulfilled the requirements of the International Protocol Revision 2010.