Report of AAPM Task Group 219 on independent calculation-based dose/MU verification for IMRT.

2021 
Independent verification of the dose per monitor unit (MU) to deliver the prescribed dose to a patient has been a mainstay of radiation oncology quality assurance. We discuss the role of secondary dose/MU calculation programs as part of a comprehensive Quality Assurance (QA) program. This report provides guidelines on calculation-based dose/MU verification for intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT) provided by various modalities. We provide a review of various algorithms for "independent/second check" of monitor unit calculations for IMRT/VMAT. The report makes recommendations on the clinical implementation of secondary dose/MU calculation programs; on commissioning and acceptance of various commercially available secondary dose/MU calculation programs; on benchmark QA and periodic quality assurance; and on clinically reasonable action levels for agreement of secondary dose/MU calculation programs.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    101
    References
    1
    Citations
    NaN
    KQI
    []