Losses of poly- and perfluoroalkyl substances to syringe filter materials

2019 
Abstract Syringe filters are used to separate solids from liquids prior to analysis of poly- and perfluoroalkyl substances (PFASs). This is a critical step in sample preparation, as losses of PFASs to the filter material can be significant and lead to underestimation. This study evaluated losses of 21 PFASs in three different matrices (methanol, MilliQ water, and water containing 10 mg L−1 dissolved organic carbon (DOC)) to six different types of syringe filter (0.45 and 0.22 μm). Regarding sample matrix, the lowest average ∑21PFAS losses were observed in methanol (13%), followed by DOC water (19%) and MilliQ water (26%). Regarding syringe filter material, the lowest average losses of ∑21PFAS in DOC water and MilliQ water were observed for a recycled cellulose filter (average losses 16% and 21%, respectively), while a polypropylene filter had the lowest ∑21PFAS losses in methanol (9%). A smaller polyethersulfone (PES) filter (0.22 μm, 17 mm O) showed significantly (p