Using procedural blanks to generate analyte-specific limits of detection for persistent organic pollutants based on GC-MS analysis.

2014 
Several methods are used to generate a limit of detection for organic pollutants measured by gas chromatography–mass spectroscopy (GC-MS); all have theoretical and practical drawbacks. The current project investigated two common existing methods (statistical and empirical) for applicability to chromatographic properties from real samples, comparing these with a new proposed method using procedural blanks to estimate a minimum detectable peak area. Weaknesses of all three methods are discussed. The proposed method was superior to other examined methods in that it provided analyte-specific limits of detection linked to the recovery of mass-labeled internal standards for every analyte within every sample. Other identified quality assurance benefits included the following: enhanced protection against false positives; providing a sensitivity performance metric across batch, analyst, and instrument; enabling chemists with discretionary decisions specific to every analyte regarding detectability and interference...