Computed tomography based assessment of programmable shunt valve settings
2021
Abstract Introduction Programmable shunt valve settings can sometimes be difficult to assess using classic read-out tools, warranting a skull X-ray. Research question Can we use available head computed tomography (CT) scans to determine the valve settings, in order to obviate the need for additional skull X-rays? Material and methods The valve setting of two different programmable shunts (Codman Certas Plus® and Sophysa Polaris®) were assessed by two blinded observers in 24 patients using 65 head CT scans (slice thickness ≤2 mm). Using multi-planar reconstruction (MPR) tools, images were resliced according to the direction of the valve, allowing a direct readout of the valve settings. We validated our CT based method against 32 available skull X-rays. Results For all CT scans it was possible to assess the valve setting. No interobserver variability was found and there was a 100 % concordance between the CT based method and skull X-rays. Discussion CT based assessment of programmable shunt valve settings is feasible and reliable. It may obviate the need for additional skull x-rays when a head CT scan is available. Conclusions This technique can reduce radiation exposure and can be applied to historical CT imaging with unknown valve settings.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader