Analyses Of H*(10) Dose Rates Measured in Environment Contaminated by Radioactive Caesium: Correction of Directional Dependence of Scintillation Detectors.
2021
Ambient dose equivalent rates were measured in the environment of the Fukushima prefecture using NaI(Tl)/CsI(Tl) scintillation detectors and CdZnTe/Ge semiconductor detectors. The dose rates obtained at the same locations varied beyond uncertainty (1σ). By replacing the spectrum-dose conversion operators obtained from the anterior-posterior geometry with those from the rotational geometry, the dose rates agreed with each other within uncertainties, except for a CsI(Tl) scintillation detector with a considerably flat crystal configuration, due to its excessive directional dependence.