Accurate Testing of Precision Voltage Reference by DC-AC Conversion

2020 
This paper describes an accurate and fast testing technique for a precision voltage reference. Our proposed DC-AC conversion and FFT spectrum analysis method is applied and its experiment results show that measurement accuracy within ±100μV and measurement variation (repeatability) within ±10μV are feasible.