A 2:1 AV Rhythm: An Adverse Effect of a Long AV Delay during DDI Pacing and Its Prevention by the Ventricular Intrinsic Preference Algorithm in DDD Mode

2012 
A 91-year-old woman received a dual-chamber pacemaker for sick sinus syndrome and intermittently abnormal atrioventricular (AV) conduction. The pacemaker was set in DDI mode with a 350-ms AV delay to preserve intrinsic ventricular activity. She complained of palpitation during AV sequential pacing. The electrocardiogram showed a 2:1 AV rhythm from 1:1 ventriculoatrial (VA) conduction during ventricular pacing in DDI mode with a long AV interval. After reprogramming of the pacemaker in DDD mode with a 250-ms AV interval and additional 100-ms prolongation of the AV interval by the ventricular intrinsic preference function, VA conduction disappeared and the patient's symptom were alleviated without increasing unnecessary right ventricular pacing. (PACE 2011;1–4)