Key comparison BIPM.RI(I)-K6 of the standards for absorbed dose to water at 10 g cm-2 of the NPL, United Kingdom and the BIPM in accelerator photon beams

2015 
A comparison of the dosimetry for accelerator photon beams was carried out between the National Physical Laboratory (NPL) and the Bureau International des Poids et Mesures (BIPM) from 23 September to 7 October 2014. The comparison was based on the determination of absorbed dose to water at 10 g cm-2 for three radiation qualities at the NPL. The results, reported as ratios of the NPL and the BIPM evaluations (and with the combined standard uncertainties given in parentheses), are 1.0000(62) at 6 MV, 0.9999(70) at 10 MV and 0.9993(80) at 25 MV. This result is the seventh in the on-going BIPM.RI(I)-K6 series of comparisons. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).