Position dependent spatial and spectral resolution measurement of distributed readout superconducting imaging detectors

2008 
We present direct measurements of spatial and spectral resolution of cryogenic distributed readout imaging detectors (DROIDs). The spatial and spectral resolutions have been experimentally determined by scanning a 10μm spot of monochromatic visible light across the detector. The influences of the photon energy, bias voltage, and absorber length and width on the spatial and spectral resolutions have been examined. The confinement of quasiparticles in the readout sensors (superconducting tunnel junctions) as well as the detector’s signal amplitude can be optimized by tuning the bias voltage, thereby improving both the spatial and spectral resolutions. Changing the length of the absorber affects the spatial and spectral resolutions in opposite manner, making it an important parameter to optimize the DROID for the application at hand. The results have been used to test expressions for photon energy, position, and spatial and spectral resolutions which have been derived by using an existing one-dimensional mod...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    2
    Citations
    NaN
    KQI
    []