In vitro examination of the primary stability of three press-fit acetabular cups under consideration of two different bearing couples

2019 
Abstract Backround For preclinical statements about the anchoring behavior of prostheses, the primary stability of the prosthesis is of special importance. It was the aim of this study to examine and compare the relevant relative micromotions of three different acetabulum prostheses by introducing three-dimensional torques. Methods The cups were implanted under standard conditions into an anatomical artificial bone model. Three-dimensional torques were applied to the acetabular cups. Taking into account the resulting frictional moments of two different bearing couples, ceramic-on-ceramic and ceramic-on-polyethylene, the relative micromotions of the cups were recorded as maximum total micromotion, translational and rotational micromotion, and the primary stability values of the three cups were compared. Results Relative micromotion of all cup models was always significantly smaller with the CoC bearing couples than with the CoP bearing couples ( p p p p Conclusion The results of our study can be interpreted as indicating that the low relative micromotions of all cups – irrespective of the use of CoC or CoP bearing couples – are within an acceptable range favoring secondary osseointegration of the implants. Furthermore, we were able to show that the cup wall thickness and the surface quality of the cup systems have an influence on the primary stability and the elastic deformability of the examined cup systems.