Capillary-Channeled Polymer (C-CP) Films as Processing Platforms for Protein Analysis by Matrix-Assisted Laser/Desorption Ionization Mass Spectrometry (MALDI-MS)

2012 
Polypropylene (PP) capillary-channeled polymer (C-CP) films have parallel, μm-sized channels that induce solution wicking via capillary action. Efficient mass transport from the solution phase to the channel surface leads to adsorption of hydrophobic protein solutes. The basic premise by which C-CP films can be used as media to manipulate analyte solutions (e.g., proteins in buffer), for the purpose of desalting or chromatographic separation prior to MALDI-MS analysis is presented here. Cytochrome c and myoglobin prepared in a Tris-HCl buffer, and ribonuclease A, lysozyme, and transferrin prepared in phosphate buffered saline (PBS), are used as the test solutions to demonstrate the desalting concept. Protein analysis is performed after deposition on a C-CP film with and without a water washing step, followed by spray deposition of a typical sinapinic acid matrix. Extracted MALDI mass spectra exhibit much improved signal-to-noise characteristics after water washing. A mixture of cytochrome c and myoglobin (2 μL of 2.5 μM each in Tris-HCl buffer) was applied, washed with water and spatially separated via simple capillary action (wicking) using a reversed-phase solvent composition of 0.1% trifluoroacetic acid (TFA) in 50:50 acetonitrile (ACN):H2O. Subsequent application of sinapinic acid followed by imaging of the film using MALDI-MS reveals that as the protein solution is wicked down the film, separation occurs.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    2
    Citations
    NaN
    KQI
    []