Hand-held mass spectrometer for environmentally relevant analytes using a variety of sampling and ionization methods
2010
A recently developed hand-held, rectilinear ion trap mass spectrometer, capable of performing in situ analysis, has been evaluated for a variety of environmentally relevant analytes. Different sampling and ionization methods were implemented, demonstrating the considerable versatility of this instrument. A discontinuous (viz. pulsed) atmospheric pressure inlet (DAPI) was used to introduce externally-generated analyte ions. Nitro compounds were ionized by electrosonic spray ionization (ESSI) yielding the protonated and sodiated forms of the molecular ion, as well as fragment ions. The amines 2,2,6,6-tetramethylpiperidine, triethylamine and 2,6-diphenylpyridine showed low parts per billion (ppb) detection limits. Vapor phase external ionization was used to examine the chemical warfare simulant dimethyl methylphosphonate and the insect repellant N,N-diethyl-m-toluamide. Membrane introduction mass spectrometry (MIMS) was used as the introduction system for hydrophobic analytes using a selectively permeable (p...