Sensitive and Quantitative Detection of C-Reaction Protein Based on Immunofluorescent Nanospheres Coupled with Lateral Flow Test Strip
2016
Sensitive and quantitative detection of protein biomarkers with a point-of-care (POC) assay is significant for early diagnosis, treatment, and prognosis of diseases. In this paper, a quantitative lateral flow assay with high sensitivity for protein biomarkers was established by utilizing fluorescent nanospheres (FNs) as reporters. Each fluorescent nanosphere (FN) contains 332 ± 8 CdSe/ZnS quantum dots (QDs), leading to its superstrong luminescence, 380-fold higher than that of one QD. Then a detection limit of 27.8 pM C-reaction protein (CRP) could be achieved with an immunofluorescent nanosphere (IFN)-based lateral flow test strip. The assay was 257-fold more sensitive than that with a conventional Au-based lateral flow test strip for CRP detection. Besides, the fluorescence intensity of FNs and bioactivity of IFNs were stable during 6 months of storage. Hence, the assay owns good reproducibility (intra-assay variability of 5.3% and interassay variability of 6.6%). Furthermore, other cancer biomarkers (P...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
48
References
117
Citations
NaN
KQI