Improving the Sensitivity of Solid-Contact Ion-Selective Electrodes by Using Coulometric Signal Transduction
2019
A fundamental limitation of potentiometric ion sensors is their relatively low sensitivity due to the logarithmic dependence between potential and activity. Here we address this issue by exploring a recently developed coulometric transduction method for solid-contact ion-selective electrodes (SCISEs). Spin-coated thin-layer ion-selective membranes are used to lower the membrane resistance and shorten the response time of the SCISEs. When using coulometric transduction, an optimized design of the K+-SCISE is able to detect a concentration change of 5 μM at a concentration level of 5 mM, corresponding to a 0.1% change in K+ activity. This indicates that SCISEs can provide extremely high sensitivity when employing coulometric transduction. Impedance measurements show that the coulometric transduction process for the K+-SCISE is limited by diffusion even for very thin ion-selective membranes. On the other hand, the H+-SCISE shows a low impedance and a fast coulometric response that is related to the high mobi...