A cementless, proximally fixed anatomic femoral stem induces high micromotion with nontraumatic femoral avascular necrosis: A finite element study

2014 
Summary Decrease in bone mineral density of metaphysis in patients with nontraumatic avascular necrosis of the femoral head (AVN) is considered the main factor leading to aseptic loosening of the femoral component. Researchers have hypothesized that a cementless, anatomic stem fixed proximally to the metaphysis has a higher risk for aseptic loosening than a straight stem that is fixed at the diaphysis in patients with nontraumatic AVN. The purpose of the current study was to evaluate the effects of cancellous bone stiffness at the metaphysis and stem geometry on the micromotion of the femoral stem relative to the femur. The VerSys (straight) and ABG (anatomic) femoral stems were enrolled in this finite element study to determine the performance of prosthetic micromotion. The simulated load to the hip joint during heel strike was assigned. Results showed that the VerSys model represented better resistance in micromotion between the bone/stem interface than the ABG model in either normal or poor cancellous bone stiffness at the metaphysis. The bone quality at the metaphysis of patients with nontraumatic AVN should be considered prior to selecting a femoral stem. In consideration of initial stability, acementless, straight stem that fits the isthmus is more favourable than an anatomic stem that is fixed to the proximal area of the canal.