Sequential cloud-point extraction for toxicological screening analysis of medicaments in human plasma by high pressure liquid chromatography with diode array detector

2013 
Abstract A complex extraction system with the use of cloud-point extraction technique (CPE) was developed for sequential isolation of basic and acidic/neutral medicaments from human plasma/serum, screened by HPLC/DAD method. Eight model drugs (paracetamol, promazine, chlorpromazine, amitriptyline, salicyclic acid, opipramol, alprazolam and carbamazepine) were chosen for the study of optimal CPE conditions. The CPE technique consists in partition of an aqueous sample with addition of a surfactant into two phases: micelle-rich phase with the isolated compounds and water phase containing a surfactant below the critical micellar concentration, mainly under influence of temperature change. The proposed extraction system consists of two chief steps: isolation of basic compounds (from pH 12) and then isolation of acidic/neutral compounds (from pH 6) using surfactant Triton X-114 as the extraction medium. Extraction recovery varied from 25.2 to 107.9% with intra-day and inter-day precision (RSD %) ranged 0.88–1087 and 5.32–17.96, respectively. The limits of detection for the studied medicaments at λ 254 nm corresponded to therapeutic or low toxic plasma concentration levels. Usefulness of the proposed CPE-HPLC/DAD method for toxicological drug screening was tested via its application to analysis of two serum samples taken from patients suspected of drug overdosing.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    18
    Citations
    NaN
    KQI
    []