SU-E-T-255: Optimized Supine Craniospinal Irradiation with Image-Guided and Field Matched Beams
2014
Purpose: Conventional craniospinal irradiation (CSI) challenges include dose inhomogeneity at field junctions and position uncertainty due to the field divergence, particular for the two spinal fields. Here we outline a new supine CSI technique to address these difficulties. Methods: Patient was simulated in supine position. The cranial fields had isocenter at C2/C3 vertebral and were matched with 1st spinal field. Their inferior border was chosen to avoid the shoulder, as well as chin from the 1st spine field. Their collimator angles were dependent on asymmetry jaw setting of the 1st spinal field. With couch rotation, the spinal field gantry angles were adjusted to ensure, the inferior border of 1st and superior border of 2nd spinal fields were perpendicular to the table top. The radio-opaque wire position for the spinal junction was located initially by the light field from an anterior setup beam, and was finalized by the portal imaging of the 1st spinal field. With reference to the spinal junction wire, the fields were matched by positioning the isocenter of the 2nd spinal field. A formula was derived to optimize supine CSI treatment planning, by utilizing the relationship among the Yjaw setting, the spinal field gantry angles, cranial field collimator angles, and the spinal field isocenters location. The plan was delivered with portal imaging alignment for the both cranial and spinal junctions. Results: Utilizing this technique with matching beams, and conventional technique such as feathering and forwarding planning, a homogenous dose distribution was achieved throughout the entire CSI treatment volume including the spinal junction. Placing the spinal junction wire visualized in both spinal portals, allows for precise determination and verification of the appropriate match line of the spine fields. Conclusion: This technique of optimization supine CSI achieved a homogenous dose distributions and patient localization accuracy with image-guided and matched beams.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI