Rapid and reproducible infectivity end-point titration of virulent phage in a microplate system

2005 
Abstract The standard method for measuring the number of infectious phages in solution has traditionally been the plaque forming assay. An alternative method is described where the number of lytic, infectious phages is determined in an endpoint titration assay adapted for a microplate system. In this model system, susceptible Escherichia coli B6 at a density of 4 × 10 7  cells/ml, were mixed with an equal volume (100 μl) of ΦX174 diluted serially in a microtest plate. After 3 h of incubation on a microplate shaker the endpoint was determined spectrophotometrically and calculated according to the method of Reed and Muench. A well was considered positive for infection if the OD 630 -value was ≤10% compared to the OD 630 -value of the negative control of uninfected cells. ID 50 -titers were 2.5× higher than the PFU-titers (CV 15%) and the intra assay reproducibility revealed a CV of 9%. The method has several advantages as compared with the conventional PFU-titration. It is less time and material consuming with the possibility to assess several samples at the same time.