Revision of the failed pedicle screw in osteoporotic lumbar spine: biomechanical comparison of kyphoplasty versus transpedicular polymethylmethacrylate augmentation.

2012 
Abstract In this study, we aimed to compare of kyphoplasty versus transpedicular polymethylmethacrylate (PMMA) augmentation biomechanically in the revision of the failed pedicle screw in osteoporotic lumbar spine. Bone mineral density (BMD) of lumbar vertebrae collected from four bovines were measured. Each vertebra was decalcified with hydrochloric acid solution to obtain osteoporotic specimens. Primary polyaxial pedicle screws were inserted into the pedicles and pulled out until they failed. The pullout strength results of all specimens were recorded. Revision pedicle screws were randomly inserted into the same pedicles by either pedicle hole PMMA augmented (group 1) or kyphoplasty (Xvoid™) PMMA augmented pedicle screws (group 2). The pullout strength results of all specimens were re-recorded. The mean BMD significantly decreased from 1.686 ± 227.9 g/cm(2) to 1.432 ± 157.1 g/cm(2) following decalcification (p