Dependence of NaI(Tl) detector intrinsic efficiency on source-detector distance, energy and off-axis distance: Their implications for radioactivity measurements
2008
In this work the dependence of intrinsic efficiency of a NaI(Tl) detector of radius 3.82 cm and height 7.62 cm on source-detector distance (d), source-off-axis distance (d 0) and γ-photon energy have been investigated using analytical and Monte Carlo methods. The results showed that, for a given off-axis distance, there exists a value of the ratio of source-detector distance (d) to detector radius (R) where intrinsic efficiency is minimum. This d/R value at which minimum efficiency occurs approaches zero as off-axis distance increases and it is almost constant with increase in energy. In the region where d/R 10 are also compared.